Abstract

A simple and sensitive high-performance liquid chromatographic method for measuring the major vitamers of vitamin B 6, i.e. pyridoxal and pyridoxal 5′-phosphate (PLP), and 4-pyridoxic acid (4-PA) in plasma was developed. The vitamers and 4-PA from plasma were extracted with 0.8 mol/l perchloric acid. The separation by HPLC is accomplished using an ODS reversed-phase column and a mobile phase of 0.1 mol/l potassium dihydrogen phosphate containing 0.1 mol/l sodium perchlorate, 0.5 g/l sodium bisulfite adjusted to pH 3, at a flow-rate of 1.0 ml/min. The vitamers and 4-PA were eluted within 13 min and their concentration is determined with a fluorometric detector (excitation, 300 nm; emission, 400 nm). In this method, PLP in plasma can be determined with high sensitivity using derivatization with sodium bisulfite in the mobile phase.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.