Abstract

Recombinant antibodies with high isoelectric point are frequent since most of them are constructed from the same framework. Classically, cation exchange chromatography is used as a standard method for the determination of antibody charge heterogeneity. In contrast, in this study highly linear pH gradients were achieved by keeping the buffering capacity over the length of the gradient constant. The buffering compounds were selected to be unretained on the column and their respective concentration was adjusted in the start and end buffer of the pH gradient to achieve constant buffering capacity. This helps conserve linearity and stability of the gradient. The method allows quantification of charge variant distribution and the determination of chromatographic isoelectric point. To demonstrate the effectiveness of this novel method, a ProPac WCX-10 column was used to separate isoforms of trastuzumab biosimilar antibodies. Effects of pH gradient linearity and of varying the analytical amount of sample on the separation are shown.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.