Abstract

This paper describes an automated multiple-beam interferometry for measuring a calibrated step height of (27.00±3.00) nm nominally. The fringes captured from the multiple-beam Fizeau–Tolansky interferometer were thinned by using a written program to obtain accurate measurement. We claim that multiple-beam interferometry based on automatic fringe thinning process can provide a real time solution for calibrating step heights precisely and with high accuracy. The uncertainty budget of the multiple-beam interferometry method due to incomplete parallelism of the incident beam and the inhomogeneity of the reflecting layers was calculated automatically with a written ray tracing program. The uncertainty budget in multiple-beam interferometry was estimated to be of the order of 3.00nm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.