Abstract

The speed of tapping mode imaging with the atomic force microscope (AFM) has been increased by over an order of magnitude. The enhanced operation is achieved by (1) increasing the instrument’s mechanical bandwidth and (2) actively controlling the cantilever’s dynamics. The instrument’s mechanical bandwidth is increased by an order of magnitude by replacing the piezotube z-axis actuator with an integrated zinc oxide (ZnO) piezoelectric cantilever. The cantilever’s dynamics are optimized for high-speed operation by actively damping the quality factor (Q) of the cantilever. Active damping allows the amplitude of the oscillating cantilever to respond to topography changes more quickly. With these two advancements, 80μm×80 μm high-speed tapping mode images have been obtained with a scan frequency of 15 Hz. This corresponds to a tip velocity of 2.4 mm/s.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.