Abstract

The aim of this study is to develop a new method to measure one-dimensional dose distribution of high-energy electron using a miniature fiber-optic radiation sensor. The measurements are made by a thin plastic optical fiber with an organic scintillating sensor tip. The scintillating light in the visible wavelength range is guided to a silicon photodiode by plastic optical fiber in order to convert light output to electrical signal. The one-dimensional spatial dependence of elctron beam is measured by moving the sensor tip with uniform speed. It is shown that this fiber-optic radiation sensor has better spatial resolution than conventional ion chamber and it needs much less time to measure one-dimensional dose distribution in the high radiation fields.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.