Abstract

Recent advances in multielement solid-state detector systems for high rate, high resolution x-ray spectroscopy at noncryogenic temperatures will be described in this paper. A 16-channel silicon detector system, designed and built at BNL, has been recently operated in the NSLS machine beam ♯X19A, showing an average energy resolution of less than 250 eV FWHM, which is adequate to discriminate the fluorescence trace element against the background of elastically scattered photons in a typical EXAFS application. A larger, 128 channel system, will soon permit a higher overall count rate: ≳106 counts per second. It is shown that, in order to achieve high resolution with a solid-state detector, special care must be spent in the detector-preamplifier assembly. A low noise detectorpreamplifier may be obtained integrating the front-end devices (an FET and/or a feedback capacitor) with the detector itself.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.