Abstract
Laser absorption spectroscopy offers the potential for fast and precise trace moisture detection in gases at atmospheric pressure with a small cross sensitivity towards other molecules. We report on the development and calibration of a fibre-coupled tunable diode laser absorption spectrometer operating in the 1 - 100 ppm humidity range. The spectrometer was tested at three European humidity standards laboratories. The performance of the spectrometer was characterized by monitoring constant water vapour concentrations over several hours, yielding a good long-term stability, reproducibility and accuracy. The standard deviations of the measured water vapour concentrations for values above 5 ppm were below ±2%. Comparisons of the instrument with chilled mirror hygrometers demonstrated its fast response time. The dependences of the signal upon the flow rate and temperature are discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.