Abstract

We describe the design, construction, alignment, and calibration of a photometric ellipsometer of the rotating-analyzer type. Data are obtained by digital sampling of the transmitted flux with an analog-to-digital converter, followed by Fourier transforming of the accumulated data with a dedicated minicomputer. With an operating mechanical rotation frequency of 74 Hz, a data acquisition cycle requires less than 7 msec. The intrinsic precision attainable is high because precision is limited only by shot noise or intrinsic source instabilities, even when relatively weak continuum lamps are used as light sources. Precision may be improved by accumulating the data for consecutive cycles at a fixed wavelength. The system allows complex reflectance ratios to be determined as continuous functions of wavelength from the near infrared to the near ultraviolet spectral range. Data reduction programs can be modified to calculate complex refractive index or dielectric function spectra, or film thicknesses and refractive indices, as well as the usual ellipsometric parameters tanpsi, cosDelta.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.