Abstract

In order to investigate how the alpha1-acid glycoprotein (AGP) concentrations of neonates change in response to surgical stress, a simple high-performance liquid chromatography (HPLC)-assay for the measurement of alpha1-acid glycoprotein levels was developed. A fraction containing alpha1-acid glycoprotein was isolated from the bulk of plasma protein by addition of 0.6M perchloric acid and was then analysed directly on a short PLRP-S 4000A reversed phase column column. The method was validated by analysis of pooled plasma from healthy adults both in comparison with a calibration curve and by standard additions. The procedure was able to isolate alpha1-acid glycoprotein rapidly (<30 min) and required only 50 microl of plasma. The mean extraction recovery was 79.1% (CV 6.4%). The within-run precision for the analysis of three replicates of quality control sample ranged from +/-1.2 to +/-3.8% and the between-run precision was +/-6.1%. The method was linear (r(2)=0.988) over a concentration range from 6 to 100.0 mg/100 ml. The AGP levels in neonatal samples ranged from 25 to 93 mg/100 ml.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.