High incidences of early aseptic loosening cases have led to numerous analyses on fixation strength of femoral component (FC). This study consists of two objectives: (1) to analyze the correlation between articulating contact area shape conformities and fixation strength, and (2) to analyze the relationship between slanting box patterns of the FC and fixation strength. Two design analyses were constructed: first, using three different types of contact congruence with radius ratios of 0.50, 0.89 and flat. Second, box patterns of the FC were anteriorly slanted at 2, 4 and 6°. The fixation strength of the FC is significantly improved (38.59%) by slanting the box pattern anteriorly as compared to the non-slanted FC design.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.