Abstract
Current approaches for accurate blood pressure determination rely predominantly on invasive techniques. High Definition Oscillometry (HDO) was evaluated as a potential non-invasive approach for accurate blood pressure recordings in cynomolgus monkeys. In conscious animals, systolic, diastolic, mean arterial blood pressure (MAP) and pulse/minute were determined 15 times within approx. 9 minutes per individual. This session was performed during 3 consecutive days. Anaesthesia induced hypotension was controlled simultaneously with HDO and telemetry as reference. Repeated measurements were highly reproducible. After procedural habituation, mean MAP was 96.2 +/- 13.7 mmHg in males and 86.9 +/- 4.3 mmHg in females. Mean intraindividual coefficients of variation ranged between 10.8% and 2.4% depending on the session and parameter. Values determined by HDO corresponded to those reported for invasive techniques. Our results demonstrate, using telemetry as reference, the accuracy of HDO-based non-invasive blood pressure measurements in macaques to detect drug-related cardiovascular changes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.