Abstract

A bi-layer device was fabricated and tested for the direct collection of electrons emitted by tritium beta decay. The sensor functions at high pressures and concentrations where previously no simple and cost effective direct measurement technique existed for tritium. A polished KOVARTM (Fe-Ni-Co alloy) rod was coated with a 1-μm thick insulating layer of alumina using electron-beam evaporation, physical vapor deposition (PVD) of alumina with oxygen dosing. The alumina deposition process was optimized to minimize pinholes and obtain a stable coating with high resistivity. The detector exhibited a nanoampere electrical response over a few decades of tritium concentration, up to pure tritium at 200 kPa. The sensor has been in service for several months now without showing signs of degradation and no discernible physical damage or change in efficiency has been observed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.