Abstract

A noncontact type centrifugal pump without any complicated control or sensing modules has been developed as a long-term implantable artificial heart. Centrifugal pumps with impellers levitated by original hydrodynamic bearings were designed and have been modified through numerical analyses and in vitro tests. The hemolysis level was reduced by changing the pressure distribution around the impeller and subsequently expanding the bearing gap. Thrombus formation in the bearing was examined with in vitro thrombogenesis tests and was reduced by changing the groove shapes to increase the bearing-gap flow to 3% of the external flow. Unnecessary vortices around the vanes were also eliminated by changing the number of vanes from four to six.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.