Abstract

The pump console used at our institute is a HL20, AB, Lund, Sweden), unit with four single modular roller pumps or a combination of three roller pumps with a centrifugal pump. The unit comes with a 24-volt inbuilt battery pack in the base of the console. The batteries consist of 230-A/hour, 12-volt, gel-filled batteries. The main input voltage to the console control module is constantly monitored. Should the line voltage fall to less than 85% of the nominal voltage (220 volts), the unit automatically switches to battery This battery system will run the complete system, all pumps, supervisory modules, monitoring, and data recording in the event of power failure. The total length of battery operation time reported by the manufactures of a complete system, one arterial at per minute (LPM) (156 rpm), three ancillary pumps (50 rpm), pump display and system monitoring is approximately1.5 hours. Battery operation time depends on the conditions of the battery and the load powered by the batteries.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.