Abstract
We developed a handheld photoacoustic signal detector for sentinel lymph node (SLN) localization. The detector consists of a solid-state dye laser and a transparent ultrasound transducer (TUT). The TUT was designed as a focused-type. The TUT had an 8 MHz center frequency with a 30% bandwidth. Using the detector, we confirmed photoacoustic signals in a tube filled with methylene blue (MB) below 10-mm thick chicken tissues. We successfully performed in vivo photoacoustic detection for MB injected SLN localization and subcutaneously injected melanoma assessments. We believe that our detector will play an effective role in SLN localization and biopsy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.