Abstract

A group-phase refractive index method was developed for improving the accuracy of a conventional two-color method used in an interferometer for length measurements. This method uses two closer wavelengths to reduce the errors caused by air turbulence and the chromatic aberration. We describe the new method and demonstrate preliminarily its effectiveness by measuring the group refractive index of air using a vacuum cell. We then compared the results with those calculated using the Edlén equation. The mean difference between the measured and calculated values was 2.15×10−6 with a standard deviation of 2.0×10−6 for the vacuum cell with a length of 184.5 mm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.