Abstract
A systematic investigation of the optimum conditions for the design and operation of a gas scintillation counter has been made. The gases studied were xenon, krypton, argon, helium, and various gas mixtures. Both a 6292 and a quartz window K1306 phototube were used with and without quaterphenyl as a wavelength shifter. Under optimum conditions an energy spread of less than 4% is obtainable for the Po210 α particles. The application of the gas scintillation counter as a fast ``slow neutron detector,'' using the reaction B10(n,α)Li7, and its usefulness as a fast fission detector have been examined and discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.