Abstract
AbstractA gas chromatographic (GC) system for the sensitive, rapid, and precise analysis of nitrous oxide (N2O) was developed. The system employs a two‐part GC column that can be separated and the first section back‐flushed to remove contaminants while sample analysis continues. Nitrous oxide was detected by an electron capture detector. The GC system was suitable for analyzing small volumes of gas samples for ambient levels of N2O at the rate of 10 samples per hour with a precision of about 1 ppb.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.