Abstract

Gamma spectrometric measurements to determine the isotopic composition and total uranium mass in UO2 pellets (D = 7.5 mm; H = 3.5 mm, ρ = 10 g/cm3) were carried out. The required efficiency curve was obtained by applying the efficiency transfer method from a calibration standard (D = 65 mm; H = 20 mm) of a slightly acidified water solution. The average isotopic composition of ten UO2 pellets was consistent with values of natural uranium given by IUPAC. The average relative bias for the 235U/238U amount ratio was −0.73% using the 1001 keV gamma line for 238U and 0.50% using the 63 keV gamma line (186 keV was always used for 235U). For the total uranium mass, the mean deviation as compared to mass determinations using a balance was 5.5% using the 1001 keV gamma line for 238U and 4.3% using the 63 keV gamma line.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.