Abstract

Application of transmission and scattering gamma photon techniques for calibration of superheated emulsion detectors used for neutron dosimetry is described. The bubbles nucleated in the detector due to neutron exposure generate detectable changes in both attenuation and scattering properties of the medium, and the magnitude of change in properties depends on population density of bubbles nucleated and in turn is proportional to neutron dose. The experimental set-up consists of (137)Cs and (241)Am sources and an HPGe detector-based gamma-ray spectrometer. An indigenously developed bubble detector and a commercially available one (BTI, Canada) are used in the present study. Theoretical models for the variation in transmitted and scattered intensities through the bubble detector as a function of neutron dose are formulated, and the experimental results obtained are found to be in good agreement with the models. In the neutron dose region studied, the transmission technique shows better sensitivity than scattering technique.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.