Abstract

High performance liquid chromatography (HPLC) with a solvent gradient and absorbance detection is one of the most widely used methods in analytical chemistry. The observed absorbance baseline is affected by the changes in the refractive index (RI) of the mobile phase. Near the limited of detection, this complicates peak quantitation. The general aspects of these RI-induced apparent absorbance effects are discussed. Two different detectors with fundamentally different optics and flow cell concepts, a variable-wavelength detector equipped with a conventional flow cell and a diode-array detector equipped with a liquid core waveguide flow cell, are compared with respect to their RI behavior. A simple method to separate static – partly unavoidable – RI effects from dynamic RI effects is presented. It is shown that the dynamic RI behavior of an absorbance detector can be well described using a single, relatively easy-to-determine metric called the G-index. The G-index is typically in the order of a few seconds and its sign depends on the optical flow cell concept.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.