Abstract
Fused-Core particles have recently been introduced as an alternative to using sub-2-microm particles in chromatographic separations. Fused-Core particles are composed of a 1.7 microm solid core surrounded by a 0.5 microm porous silica layer (d(p) = 2.7 microm) to reduce mass transfer and increase peak efficiency. The performance of two commercially available Fused-Core particles (Advanced Materials Technology Halo C18 and Supelco Ascentis Express C18) was compared with sub-2-microm particles from Waters, Agilent, and Thermo Scientific. Although the peak efficiencies were only approximately 80% of those obtained by the Waters Acquity particles, the 50% lower backpressure allowed columns to be coupled in series to increase peak efficiency to 92,750 plates. The low backpressure and high efficiencies of the Fused-Core particles offer a viable alternative to using sub-2-microm particles and very-high-pressure LC instrumentation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.