Abstract

Continuous positive airway pressure is widely used in the treatment of ARF and an evaluation of the systems is important. The authors used an artificial model to test a continuous flow system with a high compliance reservoir bag. The results confirm that the system is effective in maintaining positive pressure stability within a wide range of inspiratory peak flow rates, even when a low fresh gas flow rate is employed. Nevertheless, rebreathing of expired gases is possible and may be noticeable at high expiratory flow rates, caused by the high compliance of the reservoir bag.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.