Abstract

The authors experienced two cases of alumina prosthetic components fracture among the 2879 implanted in 10 years; these components were part of 1484 prosthetic implants. In both cases ceramic fracture was the result of a surgical technical error: in the first case a 32 mm alumina head was replaced in the same tapered neck with a 28 mm one during cup revision without removing the stem; after 9 months the new head failed. In the second case the prosthesis presented recurrent sub-dislocations causing alumina cup liner chipping. Fracture of a ceramic component is actually an episodic complication (0,08% in our experience), often caused by an uncorrect surgical technique, and shall not limit the use of this material.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.