Abstract

To report the incidence of foreign body reactions associated with placement of a polyethylene mesh implant in patients treated with trapiezectomy for trapeziometacarpal osteoarthritis. Between November 2008 and September 2012, 70 hands in 66 adults with stage IV trapeziometacarpal osteoarthritis had a trapiezectomy with interposition of a spacer made of polyethylene terephthalate mesh (Anchois Ligastic, Orthomed SA, St Jeannet, France). Out of these 70 implants, 8 implants (11%) in 8 patients (mean age, 60 y; range, 49-75 y) were removed because of persistent swelling, synovitis, and pain. The mean interval between primary and revision surgery was 14 (range, 5-27) months. Histological analysis in all cases showed a foreign body giant cell reaction. Two hands showed bone resorption or carpal bone cysts similar to silicone particle synovitis. The cysts resolved after implant removal and bone grafting. In the light of these results and the available literature, we recommend not usingthis material for interposition in the treatment of osteoarthritis of the trapeziometacarpal joint. Therapeutic IV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.