Abstract

A simple, sensitive, and rapid flow-injection method is reported for the determination of benzimidazole fungicides (BFs): fuberidazole, benomyl, and carbendazim. The method is based on enhancement of the chemiluminescence (CL) signal generated during the reaction of BFs with hydrogen peroxide and copper(II) in alkaline condition in the absence of luminescent reagents. Linear calibration graphs were obtained in the concentration range of 5–2000, 25–2000, and 25–5000 ng mL−1 for fuberidazole, benomyl, and carbendazim, respectively. The relative standard deviations (n = 4) were in the range 1.2–2.6% with limits of detection (S/N = 3) 3 ng mL−1 for fuberidazole, and 15 ng mL−1 for benomyl and carbendazim. The proposed method allowed 120 injections h−1. The effect of common anions, cations, and some pesticides on the determination of BFs was also examined. The method is applied to determine BFs in spiked natural water samples using Sep-Pak C18 cartridges for solid phase extraction and preconcentration procedure. Recoveries were in the range of 94 ± 1 to 105 ± 1%. The CL mechanism is also discussed briefly.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.