Abstract

A potentiometric determination of the total water hardness by flow-injection analysis is proposed, which uses both a Cu2+ ion-selective electrode detector and a Cu2+-ion buffer stream comprising a Cu-EDTA complex and free EDTA. This method is based on potentiometric monitoring of the change in the free Cu2+-ion concentration in the buffer with a Cu2+ ion-selective electrode, caused by the reaction of Ca2+ and Mg2+ ions with free EDTA in the buffer. Chloride interference in the measurement of the Cu2+-ion concentrations with the Cu2+ ion-selective electrode was successfully eliminated by adding triethylenetetramine to the Cu2+-ion buffer. The total hardness of seawater samples could therefore be determined at a sampling rate of 30h-1 and a relative standard deviation of 0.7%. The obtained results using the proposed method were in good accord with those obtained by conventional chelatometric titration.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.