Abstract

A simple and sensitive flow injection method with spectrophotometric detection was developed for the determination of formaldehyde. The method is based on the catalytic effect of formaldehyde on the oxidation of sulfonazo III with bromate in acidic media. The decrease in absorbance of the reaction mixture was measured at 566 nm. The calibration graph was linear in the range of 0.005 to 2.80 microg ml(-1) formaldehyde at a rate of 38 +/- 4 samples h(-1). The limit of detection was 4 ng ml(-1). The relative standard deviations for ten replicate measurements of 0.20, 0.50 and 1.00 microg ml(-1) formaldehyde were 1.3, 0.8 and 0.7%, respectively. The method was applied to the determination of formaldehyde in river water, shampoo and melamine-formaldehyde resin.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.