Abstract

A flow injection/conductometric method is proposed for determing ammonia in solutions obtained from Kjeldahl digestion. The method is based on diffusion of ammonia through a PTFE membrane from an alkaline (NaOH/EDTA) medium to a deionized water stream. The change in conductance of the deionized water stream is proportional to the ammonia concentration present in the digest. The effects of flow parameters, temperature and potential interferences are reported. Approximately 100 samples can be injected per hour; the precision is about 1%. Results for total nitrogen in vegetable tissues, animal feeds and fertilizers are in good agreement with those obtained by the usual distillation/titration method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.