Abstract

A comparison between three mathematical models frequently used in flow acoustics is presented and discussed with respect to ultrasonic flow-meter performance based on the transit-time method. The flow-meter spoolpiece geometry is assumed to be a cylindrical pipe. Semi-analytical calculations employing the Frobenius power series expansion method are shown for the cases of a constant-, linear-, parabolic-, and cubic-flow profiles although the Frobenius method presented can be applied to any smooth flow profile. It is shown that the so-called deviation of measurement, often used as a measure of the flow-meter accuracy, is strongly dependent on the acoustic mode excited and the flow profile. Furthermore, differences with respect to deviation of measurement results exist among the three mathematical models analyzed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.