Abstract Spontaneous breathing effort of a mechanically ventilated patient can seriously deteriorate the treatment outcome if it is not taken into account when choosing the appropriate settings. This paper presents an improved spontaneous breathing effort model that can be used to develop improved mechanical ventilation algorithms. Through an indicative experimental study it is shown that this model accurately describes the spontaneous breathing effort of a healthy human test subject. In comparison to the commonly used sinusoidal halfwave effort model it is shown that the proposed model improved fitting quality by a factor two.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.