Abstract
Kodak EDR 2 film was calibrated across the range of beam qualities and exposure rates typically used in our cardiac catheterisation laboratory. Its dose-response curve was successfully modelled up to its saturation point of 1 Gy. The consistency of the film's response with film batch, time between exposure and processing processor and day-to-day variations in performance, was investigated. The effects of field size, exposure rate, beam quality and use of the dynamic wedge filter were quantified. The overall uncertainty in dose was estimated to be between -20% and +40%, at 160 mGy. This uncertainty increases as the film approaches its saturation point. In addition, some unusual artefacts were observed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.