Abstract

An optimization of the PRESS sequence for magnetic resonance spectroscopy is presented to simultaneously detect the important brain metabolites of glutamate (Glu) and glutamine (Gln) at field strengths of 1.5, 3, 4.7, and 7 T. Standard, clinical examinations typically use short echo times which in general are not ideal for the separation of Glu and Gln. The optimization procedure is based on numerical product operator simulations to produce yield and overlap measurements for all possible practical choices of PRESS inter-echo timings. The simulations illustrate the substantial modulations in Glu and Gln with field strength. At all field strengths, the optimized timings demonstrate a significant reduction in overlap compared to short echo PRESS, while maintaining a high metabolite signal, with Glu and Gln yields >90% when excluding T2 relaxation losses. Minimal overlap was attained at 7 T (0.3% Gln contamination in the Glu signal), and 4.7 T (1.2%). The optimized timings were applied in vivo on healthy volunteers at field strengths of 1.5 and 4.7 T.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.