Abstract

Reusable glass dishes are recommended for use with the six-stage viable impactor for size-fractionated bioaerosol sampling. However, it is not convenient to use glass dishes because they are fragile and heavy, not to mention the time-consuming preparation process prior to bioaerosol sampling. On the other hand, disposable plastic dishes have been widely used in microbiology laboratories. However, plastic materials can retain electrostatic charges and may lead to sampling bias. The objective of this study was to evaluate the sampling bias with the use of plastic dishes when a multistage viable impactor is used for airborne fungi and bacteria sampling for field sampling. Two six-stage viable impactors were placed side-by-side 1 m apart in a 147-m3 room. One was used with plastic dishes and the other with glass dishes. Compared with the concentration data obtained with glass dishes, those collected with the plastic dishes demonstrated a significant difference for both fungi and bacteria. However, there was a...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.