Abstract

Researchers from the Savannah River Technology Center, the Center for Applied Isotope Studies (CAIS) and Sampling Systems have developed a prototype Field Deployable Tritium Analysis System (FDTAS) for near-real-time measurements of environmental levels of tritium in ground and surface water. The device consists of a modified liquid scintillation counter coupled to an automatic sampler which incorporates on-line water purification. The FDTAS has been field tested at several Savannah River Site locations and has produced results comparable to laboratory analyses for low concentrations of tritium. Figures of merit obtained in the field include an average tritium background count rate of 1.5 counts per minute (cpm), tritium detection efficiency of ≈25%, and a detection limit of <10 Bq/l for a 100 minute count.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.