Abstract
Investigations of the effects of surface roughness on the utility of grazing-angle Fourier transform infrared reflection absorption spectroscopy (IRRAS) as a method for quantifying trace contamination of metal surfaces have been extended to acetaminophen, a model active pharmaceutical agent, on 316 stainless steel. The effects are more complicated than for the surfactant sodium dodecyl sulfate (SDS) on stainless steel; they include a strong surface-finish dependence of sensitivity and nonlinear behavior at surface loadings above approximately 1-2 microg cm(-2). Using data from samples in the loading range 0-0.5 microg cm(-2), unbiased partial least squared calibrations can be readily achieved for individual surface finishes with detection limits of L(D) approximately 0.15 microg cm(-2). However, as found for SDS on stainless steel, models built using data from samples of mixed surface roughness are more problematic.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.