Abstract

For dosimetric measurements using an implantable optical fiber probe with GaN (Gallium Nitride) scintillator as radioluminescence (RL) transducer, a bi-channel method is proposed to reject the background contribution of the irradiated fiber segment. It is based on spectral differences between the narrow-band light emission from GaN and the large-band background from the irradiated optical fiber. Experimental validation of this method using 6 MV photon beam has shown that the remaining background contribution after subtraction is below 1.2% for square field sizes ranging from 3 cm to 20 cm. Furthermore, a compensation method for the over-response of GaN is also proposed, since GaN is not tissue equivalent. The over-response factor of GaN exhibits a linear increase with square field aperture and depends on depth from phantom surface. This behaviour is modelled to allow compensation in specific conditions. The proposed method has been evaluated and has shown a maximum deviation of 3% for a 6 MV photon beam and 1% for an 18 MV photon beam at a depth beyond the build-up region.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.