Abstract

Improved imaging techniques and modern radiotherapy treatment delivery in the treatment field are reduced to the precise size of the tumor, which necessitates the need for small-field dosimetry. Dosimetry in small-field dosimetry is challenging because most of the available code of practice for dosimetry is based on the cavity theory concept. Some small-sized detectors show good spatial resolution and sensitivity. Of the available small detectors, the diamond detector's performance is remarkably good. Most of the centers for radiotherapy lack diamond detectors. In this situation, if a diode detector is available, we can use it for small-field dosimetry by applying the Daisy Chaining method correction methods. In this study, the diode detector's response is not over-responding because of the defective diode. So this diode cannot be used for further measurements, and we have to regularly check the performance of the diode before using it for measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.