Abstract

MRI simulation images quality of head and neck coil scanning is better than that of radiotherapy surface coil, but currently the head and neck coil is not compatible with radiotherapy positioning devices. In this paper, a new fixation device is developed based on computer reverse engineering technology, which can be used in combination with head and neck coil. This article focuses on discussing the feasibility of the new device in radiotherapy. The obtained ACR phantom and Cat phantom 504 images were used to analyze MR and CT images quality assurance indicators. The dose attenuation of 6 MV photons was measured using the ionization chamber. The results showed each index met the clinical application requirements of intracranial tumor radiotherapy, thereby it can be used in intracranial tumor radiotherapy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.