Motion analysis is employed to assess minimally invasive surgical psychomotor skills in box trainers. Tracking of laparoscopic instruments requires sensor-based systems that can be expensive, limit movements and modify their ergonomic properties. We evaluate the feasibility of using Leap Motion as a cheap, unobtrusive alternative. Four experiments were performed to determine its precision while tracking a laparoscopic instrument inside and outside a box trainer. Static long and short term precision of the Leap Motion was <2.5mm. Precision between 12 different positions within the box trainer was <0.7mm for all measured distances between neighbors. Dynamic precision when moving the instrument for 200mm ranged between 2 and 15mm. Leap Motion presents acceptable precision values inside a box trainer. Improvements are still required (e.g.: multiple instruments’ tracking). Once solved, a validation study should verify the usefulness of Leap Motion to objectively measure skills of novices and residents during training.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.