Abstract

Motion analysis is employed to assess minimally invasive surgical psychomotor skills in box trainers. Tracking of laparoscopic instruments requires sensor-based systems that can be expensive, limit movements and modify their ergonomic properties. We evaluate the feasibility of using Leap Motion as a cheap, unobtrusive alternative. Four experiments were performed to determine its precision while tracking a laparoscopic instrument inside and outside a box trainer. Static long and short term precision of the Leap Motion was <2.5mm. Precision between 12 different positions within the box trainer was <0.7mm for all measured distances between neighbors. Dynamic precision when moving the instrument for 200mm ranged between 2 and 15mm. Leap Motion presents acceptable precision values inside a box trainer. Improvements are still required (e.g.: multiple instruments’ tracking). Once solved, a validation study should verify the usefulness of Leap Motion to objectively measure skills of novices and residents during training.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.