Abstract
A new and fast high-performance liquid chromatography (HPLC) method using technology of fused-core columns for separation of fenoxycarb and cis-, trans-permethrin has been developed and used for their determination in antiparasitic veterinary shampoo. Separation of insecticides and internal standard sudan II was achieved on the fused-core column Ascentis Express RP-Amide (100 × 3.0 mm), particle size 2.7 μm, with mobile phase acetonitrile/water (55:45, v/v) at a flow rate of 1.0 mL min−1 and at temperature 60 °C. The detection wavelength of detector was set at 225 nm for both compounds and internal standard sudan II. Under the optimum chromatographic conditions standard calibration curves were measured with good linearity [r 2 = 0.99991 for fenoxycarb, r 2 = 0.99987 for trans-permethrin, and r 2 = 0.99984 for cis-permethrin (n = 8)]. Commercial samples of antiparasitic veterinary shampoo were extracted with ethanol in ultrasound bath for 5 min. A 2-μL sample volume of the filtered solution was directly injected into the HPLC system. Accuracy of the method defined as a mean recovery of insecticides from shampoo matrix was in the range 100.43–103.85 % for both insecticides.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.