Abstract
A purpose-made set-up featuring an automated fast injector allowed the easy optimization of the injected amount and the adjustment of the separation length of conventional capillaries from a minimum of 5 cm upward. It was found that a compromise in capillary length for separation efficiency and analysis time also has to take into account the injected amount, which in turn affects the sensitivity and hence the detection limit. The versatility of the system was demonstrated by the analysis of the major cations and anions in natural water samples in less than 1 min, the concurrent determination of a mixture of amino acids and carbohydrates in 160 s, and of three active ingredients in a pharmaceutical preparation in 40 s. Plate numbers were typically around 50,000 and detection limits down to 1 muM could be achieved.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.