Abstract
We present a case of a 39-year-old woman presenting to the emergency department with persistent vaginal bleeding with myoma and endometrium thickness. The qualitative urine human chorionic gonadotropin (hCG) showed positive result, however, the quantitative serum hCG had negative result. The negative serum hCG result suggests that the false-positive result was not caused by elevated circulating hCG. According to the urine hCG one-step pregnancy device, 1 mg/dL of hemoglobin may not interference the pregnancy result. Nevertheless, we found that the hemoglobin level was 40 mg/dL in the urine specimen. We designed a in vitro experiment to evaluate the effects of hemoglobin on urine pregnancy test. The concentrations of hemoglobin from 5 to 500 mg/dL did not yield a positive urine pregnancy test. On the basis of our findings, the false positive pregnancy test was not caused by hemoglobin. It is important to confirm a suspected false-positive urine hCG test using a quantitative serum hCG test. Although it is not certain the mechanism for false positive reaction in this peculiar sample, the ACON urine hCG one-step pregnancy will occasionally yield a false-positive result in this class of patients.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.