Abstract
Objectives. The purpose of this study was to determine if ethylene oxide could sterilize extracted human teeth to be used in research. Methods. An occlusal preparation was cut in freshly extracted molars, and a small hole was drilled into the pulp chamber. A suspension of Bacillus subtilis (globigii) endospores, the standard biological monitor used for ethylene oxide sterilization, was injected into the pulp cavity, and the pulp cavity access was filled with composite material and sealed with a light-cured sealant. The teeth were exposed to either a 30°C or 63°C ethylene oxide sterilization process. Following exposure, the teeth were aseptically split and cultured to reveal viable spores. Results. Sixty-four percent of the teeth exposed to “cold” ethylene oxide treatment and 80% of the teeth exposed to the “warm” treatment still contained viable spores. Significance. Ethylene oxide cannot be relied on to sterilize extracted human teeth. Therefore, before they are used in research, other methods should be used to ensure killing of bloodborne pathogens that may be present within the teeth.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.