Abstract
This series of FactFinders presents a brief summary of the evidence and outlines recommendations regarding the use of motor stimulation testing in lumbar radiofrequency neurotomy and performance of radiofrequency neurotomy in patients with posterior spinal hardware.The evidence in support of the following facts is presented: (1) Motor stimulation does not inherently protect against unwanted damage to the spinal nerve, exiting spinal nerve root or its ventral ramus due to a lack of sensitivity of this test for identification of electrode contact or close proximity to sensorimotor nerves. Even when motor stimulation is performed, verification of correct electrode placement with multiplanar imaging including a minimum of true anterior-posterior and lateral fluoroscopic views is a recommended safeguard. (2) The existence of posterior spinal hardware is not an absolute contraindication to radiofrequency neurotomy, but direct contact with hardware should be avoided.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.