Abstract

A method of achieving a pre-stressed pressure sensitive diaphragm is discussed in this paper. A design is developed, implemented and demonstrated for a Fabry-Perot interferometer (FPI) based pressure sensor of range 0-1000 mbar. The diaphragm is designed as a ridged structure made of SS316L. It is thermally expanded and mounted on flange (sensor body). During cooling, diaphragm’s contraction gets ceased by the flange, resulting in pre-stressed diaphragm. The amount of pre-stress is governed by diametric deficit between the diaphragm and the flange. The developed sensor is tested using white light interferometry scheme and experimental results are compared with the finite element analysis results. The repeatability and accuracy are found to be 1.22% and 1.56% of full scale, respectively. To the best of our knowledge, such design and method of achieving pre-stressed diaphragm in FPI pressure sensor is reported for the first time.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.