Abstract
This paper reports on a new lithographic process for fabricating arrays of tall, high aspect ratio (defined as height/wall thickness), hollow, polymer microneedles on a platform. The microneedles feature a high sharpness (down to 3 µm tip radius) and aspect ratio (>65) which is a factor 2 and 4 better than the state of the art, respectively. The maximum achievable needle shaft length is over 1 mm. The improved performance was obtained by using an anisotropically patterned silicon substrate covered with an antireflective layer as mold for the needle tip and an optimized SU-8 lithographic process. Furthermore, a platform containing liquid feedthroughs holding an arbitrary number of needles out of plane can be manufactured with only one additional process step. The high aspect ratio microneedles undergo failure at the critical load of around 230 mN in the case of 1 mm long hollow needles with triangular cross section and a base of 175 µm. Penetration into human skin is demonstrated as well.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.