Abstract
A prototype microfluidic device using boron-doped diamond (BDD) electrodes patterned on an alumina chip was designed and fabricated. Electrochemical microfluidic devices have advantages in that the amount of sample required is small, the measurement throughput is high, different functions can be integrated on a single device, and they are highly durable. In using the device for the flow injection analysis of oxalic acid, the application of a brief conditioning step ensured that the reproducibility of the current signal was excellent.Furthermore, the fabricated system also performed as a prototype of “elimination-detection flow system”, in which interfering species are eliminated using “elimination electrodes” prior to the species reaching the “detection electrode”. The fabricated device reduced the current due to interfering species by 78%. Designs of devices to improve this efficiency are also discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.