Abstract
Abstract. The electrical conductivity (EC) of a medium invaded by TDR sensors can be estimated from the impedance of a TDR reflectogram. Four categories of sensor were tested in salt solutions and the impedances of the TDR pulse wave were correlated to the EC of the solution. The relation between the impedance and EC over a wide range of conductivities is non-linear but stable. Second- to fourth-degree polynomials can extend the measurement range to 44 dS m-1 (equivalent to a NaCl concentration of 28 g l-1 or 0.48 N) and result in better prediction of the conductivities than linear relations. For automatic measurement of EC with a datalogger, the method of Giese and Tiemann (1975, Adv. Mol. Rel. Processes, 7: 45-59) gives accurate measurement of conductivities lower than 10 dS m-1. Polynomial relations between EC and the datalogger's record provide an accurate estimate of the conductivity over a wide range. However, for both manual and automatic measurements, the sensors need to be calibrated individually. In particular, in the non-linear region, the differences between sensors are larger. Fortunately, the relation is sufficiently stable to eliminate significant error.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.