Abstract

Pulsed Field Gradient (PFG) NMR is recognised as an analytical technique used to characterise the tortuosity of porous media by measurement of the self-diffusion coefficient of a fluid contained within the pore space of the material of interest. Such measurements are usually performed on high magnetic field NMR hardware (>300 MHz). However, many materials of interest, in particular heterogeneous catalysts, contain significant amounts of paramagnetic species, which make such measurements impossible due to their characteristic short spin-spin relaxation times. Here it is demonstrated that by performing PFG NMR measurements on a low field magnet (2 MHz), tortuosity measurements can be obtained for a range of titania (TiO2) based carriers and catalyst precursors containing paramagnetic species up to a 20 wt.% loading. The approach is also used to compare the tortuosity of two catalyst precursors of the same metal loading prepared by different methods.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.